PROVA 4 to 20 mA Loop Calibrator
FEATURES:
4-20mA, with 1 p A resolution
High Accuracy 0.025% + 5 p A
Auto-Ramp and Easy Step Functions
1-100% Input
0-20mA, 0-24mA Selectable
Easy Just-Type-in Number Operation
SPECIFICATIONS: (23°C ± 5°C)
DC Current: (1KΩ maximum load, 24V Loop Supply)
| Range | Resolution | Accuracy |
| 0 to 4mA | 1μA | ± 0.025% ± 10μA |
| 4 to 20mA | 1μA | ± 0.025% ± 5μA |
| 20 to 24mA | 1μA | ± 0.025% ± 5μA |
DC Voltage:
| Range | Resolution | Accuracy |
| 0 to 4V | 1mV | ± 0.05% ± 10mV |
| 4 to 20V | 1mV | ± 0.05% ± 5mV |
| 20 to 24V | 1mV | ± 0.05% ± 5mV |
Beeper warning when output is short and specified output > 100mV
General specifications:
| Dual Display | 5 digits for the main (large) display 4 digits for the upper (small) display |
| Battery type | Alkaline battery (9V) |
| Operating temperature | 0°C to 50°C (32°F to 122°F) |
| Dimensions | 88 x 168 x 26mm (3.46" x 6.61" x 1.03") |
| Weight | 330g (batteries included), 12.34oz |
| Accessories included | Instruction Manual, 9V Battery, Carrying Case, External Battery Pack, Alligator Clip |
PROVA 4 to 20 mA Loop Calibrator
SKU: PR-100